Zebrafish exhibit a light avoidance response that can pose additional challenges for fluorescence imaging. Standard fluorescence microscopy delivers excitation light through the objective lens. At magnifications greater than approximately 2.5x this will result in non-uniform illumination of the SideView™ well. The zebrafish will tend to move away from the focal point to the dimmer regions of the well. As the objective power is increased, the intensity of the illumination beam is increased and the animals light avoidance response is further aggravated. Even at low magnification when the well is uniformly illuminated, introduction of light through the side of the well may cause some of the animals to roll onto their sides.
The solution is to flood the SideView™ wells will excitation light from above. PSI has used this approach with great success.
An LED-based illuminator similar to the one depicted below may be ordered on a custom basis.

Top illumination floods entire well with excitation source.




Epi-Illumination
Top-Illumination